CRY2304 Noise and vibration analysis system

CRY2304 Noise and vibration analysis system

CRY2304 Noise and vibration analyzer is a new type of test instrument for noise and vibration analysis launched by CRYSOUND. The host uses USB 2.0 transmission protocol to communicate with the host computer, and supports synchronous input of noise and vibration signals with 4 channels.

The analysis frequency range is 10 Hz - 80 kHz, and its noise floor is as low as -110 dBV, which is sufficient for high precision acquisition and high-quality analysis of noise and vibration in most cases.

  • Comprehensive interfaces.
  • Versatile functionality.
  • Precise measurement.

Comprehensive interface and functionality

The CRY2304 is a high-performance data acquisition and signal analysis system. It integrates multiple analog input channels with built-in ICP power supply for simultaneous noise and vibration measurement.

The device supports RS-485 communication and Modbus RTU protocol, enabling remote instrument control, data retrieval, and automated measurement processes.

Bundled CRY2300 software supports time-domain, octave, and FFT analysis on Windows 7/10, meeting diverse data processing and analysis needs.

Sleek interface.
user-friendly operation.
Simultaneous support for noise and vibration testing.
Comprehensive interface and functionality

The CRY2304 is a high-performance data acquisition and signal analysis system. It integrates multiple analog input channels with built-in ICP power supply for simultaneous noise and vibration measurement.

The device supports RS-485 communication and Modbus RTU protocol, enabling remote instrument control, data retrieval, and automated measurement processes.

Bundled CRY2300 software supports time-domain, octave, and FFT analysis on Windows 7/10, meeting diverse data processing and analysis needs.

Sleek interface.
user-friendly operation.
Simultaneous support for noise and vibration testing.

Technical Specifications

No items found.
No items found.
Please choose a topic.
Technical specification
Number of acquisition channels
Number of acquisition channels
4-channel input
ADC sampling resolution
ADC sampling resolution
24-bit
ADC sampling frequency
ADC sampling frequency
44.1kHz~192kHz
Test frequency range
Test frequency range
10Hz~80kHz
ICP /lEPE power supply
ICP /lEPE power supply
4mA(24VDC)
Input impedance
Input impedance
10kΩ
Frequency Response
Frequency Response
<±0.1dB(20Hz~20kHz)
Gain
Gain
-20dB~30dB
Noise floor
Noise floor
-100dBV(Bandwidth:20kHz,A-weighted)
Total Harmonic Distortion
Total Harmonic Distortion
0.001%(1V,1000Hz)
Equipment power supply
Equipment power supply
50Hz,220V+10%
Ambient conditions for operation
Operating temperature range
Operating temperature range
0~+40℃
Humidity range
Humidity range
≤80% RH
Physical
Weight
Weight
3500g
Size
Size
L*W*H=284mm*332mm*70mm

Download

Get in touch

Are you seeking more information about CRYSOUND's solutions or need a demo? Contact us via the form bleow and one of our sales or support engineers will connect with you.
Multi-channel
Sound and Vibration
Noise Monitors
Addition Plug-in
Microphone Array
Intrinsic Safe
Test Contents
Number of Channels
IEC 61672-1:2013 Class
Feature
Structure
Operation
Size of Microphone
Polarization
Contact Testing
Application